High-Temperature MoSi₂ Heating Elements (1700°C-1850°C)
Standard U-Shape Element
This is the most common configuration for lab muffle furnaces, dental sintering ovens, and general high-temperature kilns.
Helical / Spiral Heaters
This wound design fits horizontal tube furnaces and provides consistent heat distribution along the full chamber length.
Vertical Suspension L-Shape
This shape is designed for top-loading vertical furnaces, with angled terminals that support free vertical movement during thermal expansion.
Standard Configurations & Custom Engineering
Agile Dimension Procurement: For standard shapes, no engineering drawings are required. We custom-machined lengths and shank spacings according to your dimensions.
Straight & L-Shape
Specify hot zone diameter (d), terminal diameter (D), heating length (Le), cold length (Lu)
Multi-Shank W-Shape
Commonly used for horizontal furnace placements. Specify center spacing (a).
U-Shape with Terminal Bend
Elements bent at 45° or 90° at the terminals to protect connection leads from furnace heat.
Special Shapes & Chamber Cylinders
Complex non-standard geometries including coaxial heaters, cup shapes, and cylindrical modules with pre-fitted ceramic fiber linings.
Full Schematic Customization
We fabricate multi-bend custom MoSi₂ heaters directly from your technical blueprints. Every design is verified for voltage compatibility and expansion margins before production.
Preventing Joint Burnout: Micro-Plasma Sintering & High-Density SiO₂ Glaze
The transition zone between the heating section (d) and the cold terminal (D) is highly susceptible to mechanical and thermal stress. Gurdoo resolves this common failure point with a specialized micro-plasma welding process:
- Low-Resistance Transition Taper: By eliminating structural micro-voids, our welding ensures uniform current density across the joint, preventing localized overheating under high electrical loads.
- Pre-Passivated SiO₂ Protection: A dense, pre-formed quartz glaze layer completely seals the weld joint, isolating the molybdenum core from oxygen during rapid thermal cycling.
Minimizing Kiln Downtime: 24-Hour Dispatch for Standard MoSi₂ Replacements
Heating element failure shouldn’t halt your production or lab schedule. We maintain active inventory visibility across our manufacturing network for rapid aftermarket replacement:
- In-Stock Standard Elements: Standard U-shape elements ready to ship within 3 days.
- Matching Accessories: ension clamps, copper leads, and hanger bricks are available to bundle with element orders – no separate sourcing delay.
- Bulk Distributor Orders: Batch fulfillment available for aftermarket repair networks, with consolidated shipping to reduce freight cost.
Matching Installation Accessories
Extreme-Heat Industry Applications
Dental Zirconia Sintering
1550°C–1650°C sintering cycles for crowns and bridges.
High-Temp Testing
Furnace calibration at continuous 1800°C operation.
Semiconductor LPCVD
Clean-heat elements for high-vacuum deposition chambers.
Chamber Uniformity
Within ±5°C temperature consistency across the hot zone (dependent on furnace design).
Global Transit Protection & Double Packing
Standard Export Packing
Plywood crates with shaped EPE foam inserts. Elements are fixed in place to prevent movement during shipping.
Suspended Shockproof Packing (on request)
Elements are suspended in multi-layer EPE foam blocks. Protects bent terminal corners from vibration damage during ocean freight.
Physical Properties & Atmosphere Limit Parameters
| Volume Density (g/cm³) | Bending Strength (kg/cm²) | Vickers Hardness | Apparent Porosity | Water Absorption | Thermal Elongation |
|---|---|---|---|---|---|
| 5.5 – 7.0 | 15 – 25 | (HV) 570 kg/mm² | 7.4% | < 0.2% | 4% |
| Working Furnace Atmosphere | Max Element Operating Temperature (°C) | |||
|---|---|---|---|---|
| 1700 Type | 1800 Type | 1850 Type | 1900 Type | |
| Air | 1700°C | 1800°C | 1830°C | 1850°C |
| Nitrogen (N2) | 1600°C | 1700°C | 1700°C | 1700°C |
| Argon / Helium (Ar / He) | 1600°C | 1700°C | 1700°C | 1700°C |
| Dry Hydrogen (Dew Point -80°C) | 1150°C | 1150°C | 1150°C | 1150°C |
| Wet Hydrogen (Dew Point -20°C) | 1450°C | 1450°C | 1450°C | 1450°C |
| Other Mixed Gases (e.g., 10% CO2 / 50% CO / 15% H2) | 1600°C | 1700°C | 1700°C | 1700°C |
| Other Mixed Gases (e.g., 40% CO2 / 20% CO) | 1400°C | 1450°C | 1450°C | 1450°C |
| Partially Burnt Ammonia (8% H2) | 1400°C | 1450°C | 1450°C | 1450°C |
Precision Production via High-Tonnage CIP & Controlled Atmosphere Sintering
To maintain consistent density and electrical properties across Grade 1700, 1800 and 1900 elements, our partner facilities use high-tonnage Cold Isostatic Pressing (CIP), extrusion lines, and atmosphere-controlled sintering furnaces.
Each batch is tested for bulk density, resistance per unit length, and weld joint integrity. Test reports are available upon request.Technical Selection & Operation FAQ
Why do new MoSi₂ elements crumble at 400°C-700°C, and how can this “Pest Oxidation” be prevented?
-
How to prevent it:
- Ramp quickly through the danger zone: adjust your PID controller to ramp from 300°C to 800°C at 50°C-60°C/min. Avoid holding temperature or idling in the 400°C-700°C range.
- Pre-oxidize new elements: for new elements or furnaces after long shutdowns, heat to 1200°C-1300°C in air and hold for 2–4 hours. This forms a dense SiO₂ glaze that protects against lower-temperature attack.
Why do MoSi₂ elements keep burning out at the terminal joint, even with new replacements?
- Inconsistent welding quality: micro-voids at the weld create localized resistance, generating heat that eventually fractures the joint. Look for a supplier that uses micro-plasma welding (not standard arc welding).
- Clamps too tight or too loose: over-tightening restricts vertical expansion, causing mechanical stress at the joint. Under-tightening creates electrical arcing. Allow 2-3mm vertical clearance and torque clamps to the manufacturer’s spec.
- Oxidation of the exposed joint: if the weld area lacks a glaze layer, oxygen attacks the molybdenum core at high temperature. A pre-formed SiO₂ glaze is essential for elements operating above 1600°C.
Why do my MoSi₂ elements fail prematurely – and how can I extend their service life?
- Overheating above grade limit: even brief excursions above the rated temperature accelerate grain growth and oxidation. Install a high-limit controller with a fast-acting thermocouple to catch spikes before they damage the element.
- Mechanical stress from restricted expansion: if terminals are clamped too tight or bent during installation, thermal expansion creates internal cracks. Check that cold ends have 2-3mm vertical clearance and are perfectly aligned with the furnace terminals.
- Frequent thermal cycling: every heat-up/cool-down cycle stresses the oxide layer. If your process allows, reduce the number of on-off cycles per day, or lower the idle temperature instead of shutting down completely.
- Incorrect atmosphere for the grade: some grades perform differently in nitrogen, hydrogen, or mixed gases. Check the atmosphere table above – operating outside the recommended range accelerates oxidation or contamination.
Get a Selection & Quote Recommendation